Operating instructions

R&S ESCI MMEMory Subsystem
1166.6004.12 6.125 E-1
MMEMory:SELect[:ITEM]:NONE
This command deletes all data subsets from the list of device settings to be stored/loaded.
Example: "MMEM:SEL:NONE"
Characteristics: *RST value: --
SCPI: device-specific
Mode: all
This command is an event and therefore has no *RST value.
MMEMory:SELect[:ITEM]:DEFault
This command sets the default list of device settings to be stored/loaded.The latter includes:
current configuration of general device parameters (general setup)
current setting of the measurement hardware including markers
activated limit lines
user-defined color setting
configuration for the hardcopy output
Tracking generator settings
(only in conjunction with option Tracking Generator R&S FSP-B9 or External Generator Control
R&S FSP-B10)
Correction data for source calibration
(only in conjunction with option Tracking Generator R&S FSP-B9 or External Generator Control
R&S FSP-B10)
Trace data, non-used transducer factors/sets and non-used limit lines are not included.
Example: "MMEM:SEL:DEFault"
Characteristics: *RST value: --
SCPI: device-specific
Mode: all
This command is an event and therefore has no *RST value.
MMEMory:COMMent <string>
This command defines a comment ( max. 60 characters) for a device setting to be stored.
Example: "MMEM:COMM 'Setup for GSM measurement'"
Characteristics: *RST value: blank comment
SCPI: device-specific
Mode: all