Operating instructions

[SENSe:]ADEMod - Subsystem R&S ESCI
1166.6004.12 6.130 E-1
[SENSe:]ADEMod:AF:COUPling AC | DC
This command selects the coupling of the AF path of the analyzer.
Example: "ADEM:AF:COUP DC" 'Switches on DC coupling.
Characteristics: *RST value: AC
SCPI: device-specific
Mode: A
This command is only available with option R&S FS-K7 (FM Demodulator).
[SENSe:]ADEMod:AF:SPAN <numeric_value>
This command sets the span for result display AF spectrum.
The span is limited to half the measurement bandwidth of analog demodulation
(SENS:ADEM:BAND).
Example: "ADEM ON" 'Switches on the FM demodulator
"CALC:FEED 'XTIM:FM:AFSP' 'Switches on result display AF spectrum of
FM or
"CALC:FEED 'XTIM:RFP:AFSP''Switches on result display AF spectrum of
'RF power signal
"ADEM:BAND 5 MHz" 'Sets the measurement bandwidth to 5 MHz
"ADEM:AF:CENT 500kHz" 'Sets the AF center frequency to 500 kHz
"ADEM:AF:SPAN 200kHz" 'Sets the AF span to 200 kHz
Characterstics: *RST value: 2.5 MHz
SCPI: device-specific
Mode:FM
This command is only available with option R&S FS-K7 (FM demodulator).
[SENSe:]ADEMod:AF:SPAN:FULL
This command sets the maximum span for result display AF spectrum.
The maximum span corresponds to half the measurement bandwidth of analog demodulation
(SENS:ADEM:BAND).
Example: "ADEM ON" 'Switches on the FM demodulator
"CALC:FEED 'XTIM:FM:AFSP' 'Switches on result display AF spectrum of
'FM or
"CALC:FEED 'XTIM:RFP:AFSP' 'Switches on result display AF spectrum of
'RF power signals
"ADEM:BAND 5 MHz" 'Sets the measurement bandwidth to 5 MHz
"ADEM:AF:SPAN:FULL" 'Sets the AF span to 2.5 MHz
Characteristics: *RST value: -
SCPI: device-specific
Mode:FM
This command is only available with option R&S FS-K7 (FM demodulator).