Operating instructions

[SENSe:]ADEMod - Subsystem R&S ESCI
1166.6004.12 6.134 E-1
[SENSe:]ADEMod:SRATe?
This command returns the sample rate set up for the current analog demodulation measurement.
Example: "ADEM:SRAT?" 'Returns the current sample rate.
Characteristics: *RST value: -
SCPI: device-specific
Mode:FM
This command is only available with option R&S FS-K7 (FM Demodulator).
[SENSe:]ADEMod:AM[:TDOMain][:TYPE] <result type 1>,<result type 2>,<result type 3>
This command selects the result types to be created in parallel by AM demodulation.
Parameters: <result type 1/2/3>: WRITe The current measurement results will be obtained
AVERage The measurement results will be averaged over the
given # of measurements
MAXHold The maximum result values will be obtained over the
given # of measurements
MINHold The minimum result values will be obtained over the
given # of measurements
VIEW The measurement results are frozen and displayed,
i.e. they are not calculated for subsequent
measurements.
OFF The result type will not be used.
Notes:
Each value besides OFF can only be assigned to one result type at a time.
Setting all result types to OFF will deactivate the AM demodulator.
Examples: ADEM:AM AVER,MAXH,MINH 'Creates average, max hold and min hold values at
'a time
ADEM:AM WRIT,OFF,OFF 'Creates only the current measurement values
ADEM:AM OFF,OFF,OFF 'Switches AM demodulation off
Characteristics: *RST values: WRITe,OFF,OFF
SCPI: device-specific
Mode: A
This command is only available with option R&S FS-K7 (FM Demodulator).