Operating instructions

[SENSe:]ADEMod - Subsystem R&S ESCI
1166.6004.12 6.138 E-1
[SENSe:]ADEMod:FM[:TDOMain][:TYPE]<result type 1>,<result type 2>,<result type 3>
This command selects the result types to be created in parallel by FM demodulation.
Parameters: <result type 1/2/3>: WRITe The current measurement results will be obtained
AVERage The measurement results will be averaged over
the given # of measurements
MAXHold The maximum result values will be obtained over
the given # of measurements
MINHold The minimum result values will be obtained over
the given # of measurements
VIEW The measurement results are frozen and
displayed, i.e. they are not calculated for
subsequent measurements.
OFF The result type will not be used.
Notes: Each value besides OFF can only be assigned to one result type at a time.
Setting all result types to OFF will deactivate the FM demodulator.
Examples: ADEM:FMAVER,MAXH,MINH 'Creates average, max hold and min hold values
'at a time
ADEM:FMWRIT,OFF,OFF 'Only creates the current measurement values
ADEM:FMOFF,OFF,OFF 'Switches FM demodulation off
Characteristics: *RST values: WRITe,OFF,OFF
SCPI: device-specific
Mode:FM
This command is only available with option R&S FS-K7 (FM Demodulator).