Operating instructions

R&S ESCI [SENSe:]ADEMod - Subsystem
1166.6004.12 6.143 E-1
[SENSe:]ADEMod:PM[:TDOMain][:TYPE] <result type 1>,<result type 2>,<result type 3>
This command selects the result types to be measured in parallel by PM demodulation.
Parameter:
<result type 1/2/3>:
WRITe The current measured values are recorded.
AVERage The measurement results are averaged over the
given number of measurements.
MAXHold The maximum result over the given number of
measurements is obtained.
MINHold The minimum result over the given number of
measurements is obtained.
VIEW The measurement result is frozen on the display, i.e.
it is not recalculated in further measurements.
OFF The result type is not used.
Notes: Settings different from OFF can only be assigned to one result type at a time.
By setting all result types to OFF the AM demodulator is deactivated.
Examples:
ADEM:PM AVER,MAXH,MINH 'Determines average, maximum and minimum value at a
time
ADEM:PM WRIT,OFF,OFF 'Determines the current measured values only
ADEM:PM OFF,OFF,OFF 'Switches off the PM demodulator
Characteristics: *RST values: OFF,OFF,OFF
SCPI: device-specific
Mode: A
This command is only available with option R&S FS-K7 (FM demodulator).