Operating instructions

R&S ESCI [SENSe:]ADEMod - Subsystem
1166.6004.12 6.145 E-1
[SENSe:]ADEMod:PM:AFSPectrum[:TYPE] <result type 1>,<result type 2>,<result type 3>
This command selects the simultaneously measured result types of the AF spectrum of the pM
demodulated signal.
Parameter: <result type 1/2/3>: WRITe The current measurement results are recorded.
AVERage The measurement results averaged over the
given number of measurements are obtained.
MAXHold The maximum result is determined over the
given number of measurements.
MINHold The minimum result is determined over the given
number of measurements.
VIEW The measurement result in the display is frozen,
i.e. it is not recalculated in further measurements.
OFF The result type is not used.
Note:
Settings other than OFF can be assigned only to one result type at a time.
If all result types are set to OFF, the AF spectrum of the PM demodulated signal is not calculated.
The result type "AF spectrum of the PM demodulated signal" cannot be activated at the same
time as "AF spectrum of AM demodulated signal".
ADEM:PM:AFSP AVER,MAXH,MINH Determines average, maximum and minimum
'value simultaneously
ADEM:PM:AFSP WRIT,OFF,OFF 'Determines only current measurement results
ADEM:PM:AFSP OFF,OFF,OFF 'Switches calculation of AF spectrum off
Characteristics: *RST values: OFF,OFF,OFF
SCPI: device-specific
Mode: A
This command is only available with option R&S FS-K7 (FM demodulator).