Operating instructions

[SENSe:]ADEMod - Subsystem R&S ESCI
1166.6004.12 6.148 E-1
[SENSe:]ADEMod:SPECtrum:SPAN:ZOOM <numeric_value>
his command sets the frequency range for result display of RF spectrum determined from FM
demodulation data. The frequency range for result display is limited to the maximum span
(SENS:ADEM:SPEC:SPAN:MAX) or to the measurement bandwidth of analog demodulation
(SENS:ADEM:BAND).
Example:
"ADEM ON" 'Switches on the FM demodulator
"CALC:FEED 'XTIM:SPEC' 'Switches on result display RF spectrum"
"ADEM:SPEC:SPAN:MAX 5 MHz" 'Sets the maximum span to 5 MHz
"ADEM:SPEC:SPAN:ZOOM 1 MHz" 'Sets displayed span to 1 MHz
Characteristics: *RST value: 5 MHz
SCPI: device-specific
Mode:FM
This command is only available with option R&S FS-K7 (FM demodulator).
[SENSe:]ADEMod:SPECtrum[:TYPE] <result type 1>,<result type 2>,<result type 3>
This command selects the result types to be created in parallel by the RF SPECtrum measurement
with active analog demodulation.
Parameters: <result type 1/2/3>.: WRITe The current measurement results will be obtained
AVERage The measurement results will be averaged over the
given # of measurements
MAXHold The maximum result values will be obtained over the
given # of measurements
MINHold The minimum result values will be obtained over the
given # of measurements
VIEW The measurement results are frozen and displayed,
i.e. they are not calculated for subsequent
measurements.
OFF The result type will not be used.
Note: Each value besides OFF can only be assigned to one result type at a time.
Examples: ADEM:SPEC AVER,MAXH,MINH 'Creates average, max hold and min hold
'values at a time
ADEM:SPEC WRIT,OFF,OFF 'Only creates the current measurement values
ADEM:SPEC OFF,OFF,OFF 'Switches FM demodulation off
Characteristics: *RST values: OFF,OFF,OFF
SCPI: device-specific
Mode:FM
This command is only available with option R&S FS-K7 (FM Demodulator).