[SENSe:]ADEMod - Subsystem R&S ESCI
1166.6004.12 6.150 E-1
[SENSe<1|2>:]ADEMod:ZOOM[:STATe>] ON | OFF
The command enables or disables the zoom function for the FM-demodulator measurement data.
Depending on the selected measurement time and the demodulation bandwidth, the number of
recorded test points may be greater than that shown on the display.
With the zoom function enabled, exactly 501 test points of the result memory are displayed from the
time defined with [SENS:]ADEM:ZOOM:STARt.
With the zoom function disabled, data reduction is used to adapt the test points to the number of
points available on the display.
Example: "ADEM:ZOOM ON" 'Switches on the zoom function
Characteristics: *RST value: OFF
SCPI: device-specific
Mode: FM
This command is only available with option R&S FS-K7 (FM demodulator) installed
[SENSe<1|2>:]ADEMod:ZOOM:STARt 0s to measurement time
The command selects the start time for the display of individual measured values of the FM
demodulator. The maximum possible value depends on the measurement time, which is set in the
instrument and can be queried with command [SENSe]:ADEMod:MTIMe?.
If the zoom function is activated, 501 test points are displayed from the specified start time.
Example: "ADEM:ZOOM ON" 'Switches on the zoom function
"ADEM:ZOOM:STAR 500us" 'Sets the starting point of the display to 500 ยตs.
Characteristics: *RST value: 0 s
SCPI: device-specific
Mode: FM
This command is only available with option R&S FS-K7 (FM demodulator) installed