Operating instructions

R&S ESCI SENSe:CORRection Subsystem
1166.6004.12 6.157 E-1
SENSe:CORRection Subsystem
The SENSe:CORRection subsystem controls the correction of measured results by means of frequency-
dependent correction factors (e. g. for antenna or cable attenuation).
This subsystem also controls calibration and normalization during operation with the tracking generator
options (B9/B10). The measurement windows are selected by SENSe1 (screen A) and SENSe2 (screen B).
COMMAND PARAMETERS UNIT COMMENT
[SENSe<1|2>]
:CORRection
:METHod
:COLLect
[:ACQuire]
[:STATe]
:RECall
TRANsmission | REFLexion
THRough | OPEN
<Boolean>
Ttracking generator option
no query e
no query
:TRANsducer
:SELect
:UNIT
:SCALing
:COMMent
:DATA
[:STATe]
:DELete
:VIEW
:TSET
:SELect
:UNIT
:BREak
:COMMent
:RANGe<1...10>
[:STATe]
:DELete
:YIG
:TEMPerature
:AUTO
<name>
<string>
LINear|LOGarithmic
<string>
<freq> , <level> ..
<Boolean>
--
<Boolean>
<name>
<string>
<Boolean>
<string>
<freq> , <freq>, @<name> ..
<Boolean>
--
<Boolean>
HZ , --
--
HZ, HZ, --
--
no query
keine Abfrage
[SENSe<1|2>:]CORRection[:STATe] ON | OFF
This command activates/deactivates the normalization of the measurement results in the selected
window provided that the tracking generator is active. The command is available only after
acquisition of a reference trace for the selected type of measurement (transmission/reflection, see
command [SENSe<1|2>:]CORRection:COLLect[:ACQire]).
Example: "CORR ON " 'Activates normalization in screen A.
Characteristics: *RST value: OFF
SCPI: conforming
Mode: A
This command is only valid in conjunction with the tracking / ext. generator control (B9/B10).
[SENSe<1|2>:]CORRection:METHod TRANsmission | REFLection
This command selects the type of measurement with active tracking generator
(transmission/reflection).
Example: "CORR:METH TRAN " 'Sets the type of measurement in screen A to “transmission“.
Characteristics: *RST value: TRANsmission
SCPI: device specific
Mode: A
This command is only valid in conjunction with the tracking / ext. generator control option (B9/B10).