Operating instructions

R&S ESCI SENSe:CORRection Subsystem
1166.6004.12 6.159 E-1
[SENSe<1|2>:]CORRection:TRANsducer:UNIT <string>
This command defines the unit of the transducer factor selected.
Parameter: <string>::= 'DB' | 'DBM' | 'DBMV' | 'DBUV' | 'DBUV/M' | 'DBUA'
'DBUA/M' | 'DBPW' | 'DBPT'
Example: ":CORR:TRAN:UNIT 'DBUV'"
Characteristics: *RST value: 'DB'
SCPI: device-specific
Modes: A
Prior to this command, the command SENS:CORR:TRAN:SEL must be sent.
[SENSe<1|2>:]CORRection:TRANsducer:SCALing LINear | LOGarithmic
This command defines whether the frequency scaling of the transducer factor is linear or logarithmic.
Example: ":CORR:TRAN:SCAL LOG"
Characteristics: *RST value: LINear
SCPI: device-specific
Modes: A
Prior to this command, the command SENS:CORR:TRAN:SEL must be sent.
[SENSe<1|2>:]CORRection:TRANsducer:COMMent <string>
This command defines the comment for the selected transducer factor.
Example: ":CORR:TRAN:COMM 'FACTOR FOR ANTENNA'"
Characteristics: *RST value '' (empty comment)
SCPI: device specific
Modes: A
Prior to this command, the command SENS:CORR:TRAN:SEL must be sent.
[SENSe<1|2>:]CORRection:TRANsducer:DATA <freq>,<level>..
This command defines the reference values of the transducer factor selected. These values are
entered as a sequence of frequency/level pairs. The frequencies must be sent in ascending order.
Note: The level values are sent as dimensionless numbers; the unit is specified by means of the
command SENS:CORR:TRAN:UNIT.
Example: ":CORR:TRAN:TRANsducer:DATA 1MHZ,-30,2MHZ,-40"
Characteristics: *RST value: -
SCPI: device-specific
Modes: A
Prior to this command, the command SENS:CORR:TRAN:SEL must be sent.
[SENSe<1|2>:]CORRection:TRANsducer[:STATe] ON | OFF
This command switches the selected transducer factor on or off.
Example: ":CORR:TRAN ON"
Characteristics: *RST value: OFF
SCPI: device-specific
Modes: A
Prior to this command, the command SENS:CORR:TRAN:SEL must be sent.