R&S ESCI SENSe:DETector Subsystem
1166.6004.12 6.165 E-1
[SENSe<1|2>:]DETector:RECeiver[:FUNCtion] POSitive | NEGative | RMS | AVERage | QPEak
|CAVerage|CRMS
This command activates the detectors for single point measurements in EMI receiver operating
mode.
A trace cannot be selected. Three detectors can be activated at the same time.
Example: "DET:REC POS,AVER,QPE" activates the detectors Peak, Average and
Quasipeak.
Characteristics: *RST value: POS
SCPI: conforming
Mode: R
:[SENSe<1|2>:]DETector<1 to 3>:FMEasurement NEGative | POSitive | RMS | AVERage| QPEak
|CAVerage|CRMS
This command selects the detector for the final measurement (the detector used subsequent to the
prescan)
Example: ":DET:FME POS"
Characteristics: *RST value: Trace 1 QPEak
Trace 2 AVERage
Trace 3 RMS
SCPI: device-specific
Mode: R