Operating instructions

R&S ESCI SENSe:FMEasurement Subsystem
1166.6004.12 6.167 E-1
:[SENSe<1|2>:]FMEasurement:LISN[:TYPE] TWOPhase | FOURphase | OFF
This command selects the V-network to be controlled via the user port:
TWOphase = Two-line V-network ESH3-Z5
FOURphase = Four-line V-network ESH2-Z5 oder ENV4200
OFF = remote control is deactivated
Example: ":FME:LISN:TWOP"
Characteristics: *RST value: OFF
SCPI: device specific
Mode: R
:[SENSe<1|2>:]FMEasurement:LISN:PHASe L1 | L2 | L3 | N
This command selects the phase of the V-network on which the RFI voltage is to be measured.
Example: ":FME:LISN:PHAS L1"
Characteristics: *RST value: L1
SCPI: device specific
Mode: R
:[SENSe<1|2>:]FMEasurement:LISN:PEARth GROunded | FLOating
This command selects the setting of the protection earth of the V-network.
Example: ":FME:LISN:PEAR GRO"
Characteristics: *RST value: GROunded
SCPI: device specific
Mode: R
:[SENSe<1|2>:]FMEasurement:TIME <numeric_value>
This command defines the measurement time in which the values indicated in the peak list (final
measurement values) are re-checked.
Example: ":FME:TIME 0.01"
Characteristics: *RST value: 1 s
SCPI: device specific
Mode: R