R&S ESCI SENSe:FREQuency Subsystem
1166.6004.12 6.169 E-1
[SENSe<1|2>:]FREQuency:CENTer:STEP:LINK:FACTor 1 to 100 PCT
This command couples the step width of the center frequency with a factor to the span (span >0) or
to the resolution bandwidth (span = 0).
Example: "FREQ:CENT:STEP:LINK:FACT 20PCT"
Characteristics: *RST value: - (AUTO 0.1 × SPAN is switched on)
SCPI: device-specific
Mode: A
[SENSe<1|2>:]FREQuency:SPAN 0 to f
max
This command defines the frequency span in the analyzer mode.
Example: "FREQ:SPAN 10MHz"
Characteristics: *RST value: f
max
with f
max
= maximum frequency
SCPI: conforming
Mode: A
[SENSe<1|2>:]FREQuency:SPAN:FULL
This command sets the frequency span in the analyzer mode to its maximum.
Example: "FREQ:SPAN:FULL"
Characteristics: *RST value: -
SCPI: conforming
Mode: A
[SENSe<1|2>:]FREQuency:STARt 0 to f
max
This command defines the start frequency of the overall scan in receiver mode. It defines the start
frequency of the sweep in analyzer mode..
Example: "FREQ:STAR 20MHz"
Characteristics: *RST value: 0
SCPI: conforming
Mode: R, A-F
[SENSe<1|2>:]FREQuency:STOP 0 to f
max
This command defines the stop frequency of the overall scan in receiver mode. It defines the stop
frequency of the sweep in analyzer mode..
Example: "FREQ:STOP 2000MHz"
Characteristics: *RST value: f
max
SCPI: conforming
Mode: R, A-F