Operating instructions

SENSe:LIST Subsystem R&S ESCI
1166.6004.12 6.172 E-1
COMMAND PARAMETER UNIT COMMENT
[SENSe<1|2>]
:LIST
:POWer
:RESult?
[:SEQuence]
:SET
:STATe
<numeric_value>,
<numeric_value>,
<numeric_value>,
<numeric_value> | OFF,
NORMal | CFILter | RRC,
<numeric_value>,
<numeric_value>,
<numeric_value>,
<numeric_value>
...
<Boolean>,
<Boolean>,
<Boolean>,
IMMediate | EXTernal | VIDeo | IFPower,
POSitive|NEGative,
<numeric_value>,
<numeric_value>
OFF
HZ,
DBM,
DB,
DB,
--,
HZ,
HZ,
S,
PCT|DBM
--,
--,
--,
--,
--,
S,
S
Query only
[SENSe<1|2>:]LIST:POWer[:SEQuence] <analyzer freq>,<ref level>,<rf att>,OFF,
<filter type>,<rbw>,<vbw>,<meas time>,<trigger level>,...
This command configures the list of settings (max. 100 entries) for the multiple power measurement
and starts a measurement sequence. When synchronizing the command with *OPC, a service
request is generated as soon as all frequency points are processed and the defined number of
individual measurements is reached.
To reduce the setting time, all indicated parameters are set up simultaneously at each test point.
The query form of the command processes the list and immediately returns the list of results. The
number of results per test point depends on the setting of the "SENSe:LIST:POWer:SET"
command.
Parameter: Note: The following parameters are the settings for an individual frequency point.
They are repeated for every other frequency point.
For reasons of compatibility with the FSP instrument family , input OFF is
required after parameter <rf att>.
<analyzer freq>: Receive frequency for the signal to be measured
(= center frequency in manual operation)
Range of values: 0 Hz to max. frequency, depending
on the instrument model.
<ref level>: Reference level
Range of values: +30 dBm to -75 dBm in 5 dB steps
<rf att>: RF input attenuation
Range of values: 0 dB to 75 dBin 5 dB steps