Operating instructions

SENSe:LIST Subsystem R&S ESCI
1166.6004.12 6.174 E-1
Step Freq.
[MHz] Ref Level RF Att
OFF Filter type RBW VBW Meas Time TRG Level
(reserved)
1 935.2 -20 dBm 10 dB OFF Normal 1 MHz 3 MHz 434 us 0
2 935.4 -20 dBm 10 dB 10dB Channel 30 kHz 100 kHz 434 us 0
3 935.6 -20 dBm 10 dB 20dB Channel 30 kHz 100 kHz 434 us 0
"SENSe:LIST:POWer? 935.2MHz,-20dBm,10dB,OFF,NORM,1MHz,3MHz,434us,0,
935.4MHz,-20dBm,10dB,OFF,CFIL,30kHz,100kHz,434us,0,
935.6MHz,-20dBm,10dB,OFF,CFIL,30kHz,100kHz,434us,0"
performs the same measurement and returns the result list immediately after the last frequency
point.
Notes: - The measurement is performed in the time domain and therefore the span is set to 0 Hz. If
the time domain is abandoned, the function is automatically switched off.
- The measurement is not compatible with other measurements, especially as far as marker,
adjacent channel power measurement or statistics are concerned. The corresponding
commands thus automatically deactivate the function.
- The function is only available in REMOTE operation. It is deactivated when switching the
instrument back to LOCAL.
Characteristics: *RST value: --
SCPI: device-specific
Mode:A-F, A-T