Operating instructions

R&S ESCI SENSe:LIST Subsystem
1166.6004.12 6.175 E-1
[SENSe<1|2>:]LIST:POWer:SET <PEAK meas>,<RMS meas>,<AVG meas>,
<trigger mode>,<trigger slope>,<trigger offset>,<gate length>
This command defines the constant settings for the list during multiple power measurement.
Parameters <PEAK meas>, <RMS meas> and <AVG meas> define, which measurements are to be
performed at the same time at the frequency point. Correspondingly, one, two or three results per
frequency point are returned for the SENS:LIST:POW? command. If all three parameters are set to
OFF, the command generates an execution error.
Parameter:
<PEAK meas>: ON activates the measurement of the peak power (peak
detector).
OFF deactivates the measurement of the peak power.
<RMS meas>: ON activates the measurement of the RMS power (RMS
detector).
OFF deactivates the measurement of the RMS power.
<AVG meas>: ON activates the measurement of the average power (average
detector).
OFF deactivates the measurement of the average power.
<trigger mode>: Selection of the trigger source used for the list measurement.
Possible values:
IMMediate | VIDeo | EXTernal | IFPower
<trigger slope>: Used trigger slope.
Possible values:
POSitive | NEGative
<trigger offset>: Offset between the detection of the trigger signal and the start of the
measurement at the next frequency point.
Range of values: 0 s, 125 ns to 100s
<gate length>: Gate length with Gated Sweep.
Range of values: 0 s, 125 ns to 100s
Note:
The value 0 s deactivates the use of GATED TRIGGER; other
values activate the GATED TRIGGER function.
Values <> 0 s are only possible if <trigger mode> is different from
IMMediate. Otherwise, an execution error is triggered.
Returned values:
The query command returns a list of comma-separated values (CSV) of the settings, i.e.
ON,ON,ON,IMM,POS,0,0
if the configuration has been set with the command
"SENSe:LIST:POWer:SET ON,ON,ON,IMM,POS,0,0"
Examples: "SENSe:LIST:POWer:SET ON,OFF,OFF,EXT,POS,10US,434US"
"SENSe:LIST:POWer:SET ON,ON,ON,VID,NEG,10US,0"
Characteristics: *RST values: ON,OFF,OFF,IMM,POS,0S,0S
SCPI: device-specific
Mode:A-F, A-T