Operating instructions

SENSe:LIST Subsystem R&S ESCI
1166.6004.12 6.176 E-1
[SENSe<1|2>:]LIST:POWer:RESult?
This command queries the result of a previous list measurement as configured and initiated with
SENSe:LIST:POWer[:SEQuence]. The measured results are output in a list of floating point
values separated by commas. The unit of the results depends on the setting made with the
CALC:UNIT command.
This command may be used to obtain measurement results in an asynchronous way, using the
service request mechanism for synchronization with the end of the measurement.
If no measurement results are available, the command will return a query error.
Example:
'Configuration of the status reporting system for the generation of an SRQ
on operation complete
*ESE 1
*SRE 32
'Configuring and starting the measurement
"SENSe:LIST:POWer 935.2MHz,-20dBm,10dB,OFF,NORM,1MHz,3MHz,434us,0,
935.4MHz,-20dBm,10dB,OFF,NORM,30kHz,100kHz,434us,0,
935.6MHz,-20dBm,10dB,OFF,NORM,30kHz,100kHz,434us,0;
*OPC"
'Further actions of the control computer during measurement
...
'Response to service request
On SRQ:
SENSe:LIST:POWer:RESult?
Characteristics: *RST value: --
SCPI: device-specific
Mode:A-F, A-T
[SENSe<1|2>:]LIST:POWer:STATe OFF
This command deactivates the list measurement.
Example: "SENSe:LIST:POWer:STATe OFF"
Characteristics: *RST value: --
SCPI: device-specific
Mode:A-F, A-T