SENSe:MPOWer Subsystem R&S ESCI
1166.6004.12 6.180 E-1
SENSe:MPOWer:RESult[:LIST]?
This command queries the results of a multiple burst power measurement as configured and initiated
with SENSe:MPOWer[:SEQuence]. The results are output in a comma-separated list of floating
point values. The unit used for the return values is always dBm.
This command may be used to obtain measurement results in an asynchronous way using the
service request mechanism for synchronization with the end of the measurement.
If no measurement results are available, the command will return a query error.
Example: *ESE 1 'Configuration of status reporting systems for the
*SRE 32 'generation of an SRQ on operation complete
SENSe:MPOWer
935.2MHz,1MHz,434us,VIDEO,50PCT,5us,MEAN,20;*OPC
'Configuring and starting the measurement
... 'Further actions of the control computer during
'measurement
On SRQ: 'Response to service request
SENSe:MPOWer:RESult?
Characteristics: *RST value: --
SCPI: instrument-specific
Mode:A-F, A-T
SENSe:MPOWer:RESult:MIN?
This command queries the minimum power value in a multiple burst power measurement as
configured and initiated with SENSe:MPOWer[:SEQuence]. The unit used for the return values is
always dBm.
If no measurement result is available, the command will return a query error.
Example: *ESE 1 'Configuration of status reporting systems for the
*SRE 32 'generation of an SRQ on operation complete
SENSe:MPOWer
935.2MHz,1MHz,434us,VIDEO,50PCT,5us,MEAN,20;*OPC
'Configuring and starting the measurement
... 'Further actions of the control computer during
'measurement
On SRQ: 'Response to service request
SENSe:MPOWer:RESult:MIN?
Characteristics: *RST value: --
SCPI: instrument-specific
Mode:A-F, A-T