Operating instructions

SENSe:SCAN Subsystem R&S ESCI
1166.6004.12 6.188 E-1
SENSe:SCAN Subsystem
This subsystem controls the parameters for the receiver scan data. The numeric suffix in
[SENSe<1|2>]SCAN is not significant.
COMMAND PARAMETER UNIT COMMENT
[SENSe<1|2>]
:SCAN<1 to 10>
:STARt
:STOP
:STEP
:TIMe
:BANDwidth
:RESolution
:INPut
:ATTenuation
:AUTO
:GAIN
:STATe
:AUTO
:RANGes
[:COUNt]
:TDOMain
<numeric_value>
<numeric_value>
<numeric_value>
<numeric_value>
<numeric_value>
<numeric_value>
<Boolean>
<Boolean>
<Boolean>
<numeric_value>
HZ
HZ
HZ
S
HZ
DB
--
--
--
s
[SENSe<1|2>:]SCAN<1 to 10>:STARt f
min
to f
max
This command defines the start frequency of the selected receiver scan range.
Example: ":SCAN1:STAR 50kHz" 'Sets the start frequency of scan range 1 to 50 kHz
Characteristics: *RST value: 150 kHz (range1)
30 MHz (range2)
SCPI: device-specific
Mode:R
[SENSe<1|2>:]SCAN<1 to 10>:STOP f
min
to f
max
This command defines the stop frequency of the selected receiver scan range.
Example: ":SCAN1:STOP 200kHz" 'Sets the stop frequency of scan range 1 to 200 kHz
Characteristics: *RST value: 30 MHz (range1)
1 GHz (range2)
SCPI: device-specific
Mode:R
[SENSe<1|2>:]SCAN<1 to 10>:STEP f
min
to f
max
This command defines the step size for the frequency of the selected receiver scan range.
Example: ":SCAN1:STEP 100Hz" 'Sets the step size for the frequency of scan
'range 1 to 100 Hz
Characteristics: *RST value: 4 kHz (range 1)
40 kHz (range 2)
SCPI: device-specific
Mode:R