Operating instructions

R&S ESCI SENSe:SCAN Subsystem
1166.6004.12 6.189 E-1
[SENSe<1|2>:]SCAN<1 to 10>:BANDwidth:RESolution f
min
to f
max
This command defines the bandwidth for the selected receiver scan range.
Example: ":SCAN1:BAND:RES 1MHz" 'Sets the resolution bandwidth of scan
'range 1 to 1 MHz
Characteristics: *RST value: 9 kHz (range 1)
120 kHz (range 2)
SCPI: device-specific
Mode:R
[SENSe<1|2>:]SCAN<1 to 10>:TIME 100 µs to 100 s
This command defines the measurement time of the receiver subscan.
Example: ":SCAN1:TIME 1 ms" 'Sets the measurement time of scan
'range 1 to 1 ms
Characteristics: *RST value: 1 ms (range 1)
100 µs (range 2)
SCPI: device-specific
Mode:R
[SENSe<1|2>:]SCAN<1 to 10>:INPut:ATTenuation dB
min
to dB
max
This command defines the RF attenuation for the selected receiver scan range.
Example: ":SCAN1:INP:ATT 30dB" 'Sets the input attenuation of scan
'range 1 to 1 ms
Characteristics: *RST value: 10 dB
SCPI: device-specific
Mode:R
[SENSe<1|2>:]SCAN<1 to 10>:INPut:ATTenuation:AUTO ON | OFF
This command switches on or off the autoranging function in the selected receiver scan range.
Example: ":SCAN1:INP:ATT:AUTO ON" 'Activates the auto ranging function for scan
range 1
Characteristics: *RST value: OFF
SCPI: device-specific
Mode:R
[SENSe<1|2>:]SCAN<1 to 10>:INPut:GAIN:STATE ON | OFF
This command switches on or off the preamplifier in the selected receiver scan range.
Example: ":SCAN1:INP:GAIN:STAT ON" 'Activates the preamplifier for scan range 1
Characteristics: *RST value: OFF
SCPI: device-specific
Mode:R