Operating instructions

SENSe:SWEep Subsystem R&S ESCI
1166.6004.12 6.194 E-1
[SENSe<1|2>:]SWEep:POINts 125 to 8001
This command defines the number of measurement points for one sweep run.
Parameter: <numeric_value>::= 125, 251, 501, 1001, 2001, 4001, 8001
Example: "SWE:POIN 251"
Characteristics: *RST value: 501
SCPI: conforming
Modes:A
[SENSe<1|2>:]SWEep:SPACing LINear | LOGarithmic | AUTO
This command switches between linear, logarthmic and automatically selected linear step size in the
receiver. In the receiver mode, this parameter has no impact on the graphical representation of the
frequency axis.
In the analyzer, the command toggles between linear and logarithmic sweep. The frequency axis is
represented linearly or logarithmically depending on the sweep.
Example: "SWE:SPAC LOG"
Characteristics: *RST value: LIN
SCPI: conforming
Modes: R, A, VA
The frequency axis is set to linear or logarithmic scaling accordingly.