Operating instructions

SOURce Subsystem R&S ESCI
1166.6004.12 6.198 E-1
SOURce<1|2>:POWer[:LEVel][:IMMediate]:OFFSet -200dB to +200dB
This command defines a level offset for the tracking generator level. Thus, for example, attenuators
or amplifiers at the output of the tracking generator can be taken into account for the setting.
This command is only valid in conjunction with the tracking generator option R&S FSP-B9.
Example: "SOUR:POW:OFFS -10dB"'Sets the level offset of the tracking generator in
'screen A to - 20 dBm.
Characteristics: *RST- value: 0dB
SCPI: conforming