Operating instructions

SOURce Subsystem R&S ESCI
1166.6004.12 6.200 E-1
SOURce<1|2>:EXTernal<1|2>:FREQuency[:FACTor]:DENominator <numeric_value>
This command defines the denominator of the factor with which the analyzer frequency is multiplied
in order to obtain the transmit frequency of the selected generator 1 or 2 in the selected window.
Note: Select the multiplication factor in a way that the frequency range of the generator is not
exceeded by the following formula
OffsetAnalyzerGenerator
F
rDenominato
Numerator
FF += *
if applied to the start and stop frequency of the analyzer.
This command is only valid in combination with option Ext. Generator Control B10.
Example: "SOUR:EXT:FREQ:NUM 4"
"SOUR:EXT:FREQ:DEN 3" 'Sets a multiplication factor of 4/3, i.e.
the 'transmit frequency of the generator
is '4/3 times the analyzer frequency.
Characteristics: *RST value: 1
SCPI: device-specific
Mode:all
SOURce<1|2>:EXTernal<1|2>:FREQuency[:FACTor]:NUMerator <numeric_value>
This command defines the numerator of the factor with which the analyzer frequency is multiplied to
obtain the transmit frequency of the selected generator 1 or 2 in the selected window.
Note: Select the multiplication factor so that the frequency range of the generator is not exceeded
if the following formula
OffsetAnalyzerGenerator
F
rDenominato
Numerator
FF += *
is applied to the start and stop frequency of the analyzer.
This command is only valid in combination with option Ext. Generator Control B10.
Example: "SOUR:EXT:FREQ:NUM 4"
"SOUR:EXT:FREQ:DEN 3" 'Sets a multiplication factor of 4/3, i.e.
the 'transmit frequency of the generator
is '4/3 times the analyzer frequency.
Characteristics: *RST value: 1
SCPI: device-specific
Mode:all