Operating instructions

SYSTem Subsystem R&S ESCI
1166.6004.12 6.216 E-1
Notes: Generators with TTL interface can also be operated via IECBUS (= GPIB) alone.
With NONE selected, the corresponding generator 1 or 2 is deactivated.
The command is only available with option Ext. Generator Control B10.
Example: "SYST:COMM:RDEV:GEN2:TYPE 'SME02'" 'Selects SME02 as generator 2.
Characteristics: *RST value: NONE
SCPI: device-specific
SYSTem:COMMunicate:SERial:CONTrol:DTR IBFull | OFF
SYSTem:COMMunicate:SERial:CONTrol:RTS IBFull | OFF
These commands switch the hardware handshake procedure for the serial interface off (OFF) or on
(IBFull).
The two commands are equivalent.
Examples: "SYST:COMM:SER:CONT:DTR OFF"
"SYST:COMM:SER:CONT:RTS IBF"
Characteristics: *RST value: -- (no influence on this parameter, factory default OFF)
SCPI: conforming
Mode: all
SYSTem:COMMunicate:SERial[:RECeive]:BAUD 110 | 300 | 600 | 1200 | 2400 | 9600 | 19200
This command sets the transmission speed for the serial interface (COM).
Example: "SYST:COMM:SER:BAUD 2400"
Characteristics: *RST value: -- (no influence on this parameter, factory default 9600)
SCPI: conforming
Mode: all
SYSTem:COMMunicate:SERial[:RECeive]:BITS 7 | 8
This command defines the number of data bits per data word for the serial interface (COM).
Example: "SYST:COMM:SER:BITS 7"
Characteristics: *RST value: -- (no influence on this parameter, factory default 8)
SCPI: conforming
Mode: all
SYSTem:COMMunicate:SERial[:RECeive]:PARity[:TYPE] EVEN | ODD | NONE
This command defines the parity check for the serial interface (COM).
Possible values are: EVEN even parity
ODD odd parity
NONE no parity check.
Example: "SYST:COMM:SER:PAR EVEN"
Characteristics: *RST value: -- (no influence on this parameter, factory default NONE)
SCPI: conforming
Mode: all