Operating instructions

R&S ESCI SYSTem Subsystem
1166.6004.12 6.217 E-1
SYSTem:COMMunicate:SERial[:RECeive]:SBITs 1|2
This command defines the number of stop bits per data word for the serial interface (COM).
Example: "SYST:COMM:SER:SBITs 2"
Characteristics: *RST value: -- (no influence on this parameter, factory default 1)
SCPI: conforming
Mode: all
SYSTem:COMMunicate:SERial[:RECeive]:PACE XON | NONE
This command switches on or off the software handshake for the serial interface.
Example: "SYST:COMM:SER:PACE XON"
Characteristics: *RST value: -- (no influence on this parameter, factory default NONE)
SCPI: conforming
Mode: all
SYSTem:COMMunicate:PRINter:ENUMerate:FIRSt?
This command queries the name of the first printer (in the list of printers) available under Windows NT.
The names of other installed printers can be queried with command SYSTem:COMMunicate:
PRINter:ENUMerate:NEXT?.
If no printer is configured an empty string is output.
Example: "SYST:COMM:PRIN:ENUM:FIRS?"
Characteristics: *RST value: NONE
SCPI: device-specific
Mode: all
SYSTem:COMMunicate:PRINter:ENUMerate:NEXT?
This command queries the name of the next printer installed under Windows NT.
The command
SYSTem:COMMunicate:PRINter:ENUMerate:FIRSt?
should be sent previously to return to the beginning of the printer list and query the name of the first
printer.
The names of other printers can then be queried with NEXT?. After all available printer names have
been output, an empty string enclosed by quotation marks (") is output for the next query. Further
queries are answered by a Query Error.
Example: "SYST:COMM:PRIN:ENUM:NEXT?"
Characteristics: *RST value: NONE
SCPI: device-specific
Mode: all