Operating instructions

TRACe Subsystem R&S ESCI
1166.6004.12 6.230 E-1
TRACe<1|2>:IQ:SET <filter type>,<rbw>,<sample rate>,<trigger source>,<trigger slope>,
<pretrigger samples>,<# of samples>
This command defines the settings of the analyzer hardware for the measurement of I/Q data.
This allows setting the bandwidth of the analog filters in front of the A/D converter as well as setting
the sample rate, trigger conditions and the record length.
Note: If this command is omitted, the current analyzer settings will be used for the corresponding
parameters.
Parameter:
<filter type>: NORMAL 'Selects the analog analyzer resolution filters as filter type. This is
currently the only available filter type.
<rbw>: Bandwidth of the analog filters in front of the A/D converter.
<sample rate>: Sampling rate for the data acquisition.
Value range: 15.625 kHz, 31.25 kHz, 62.5 kHz,
125 kHz, 250 kHz, 500 kHz,
1 MHz, 2 MHz, 4 MHz, 8 MHz, 16 MHz,
32 MHz for <filter type> = NORMal
<trigger mode>: Selection of the trigger source used for the measurement.
Values: IMMediate | EXTernal | IFPower | RFPower
Notes:
IFPower and RFPower are available as of model 03
of the detector board assembly.
RFPower requires the TV and RF Trigger option R&S FSP-B6.
After selecting IFPower and RFPower, the trigger threshold can be set with
command TRIG:LEV:IFP or TRIG:LEV:RFP.
<trigger slope>: Used trigger slope.
Values: POSitive (currently the only value supported)
<pretrigger samples>: Number of measurement values to be recorded before the trigger point.
Range: -16744447
(= -(2
24
-1-512k)) to 65023 (= 64*1024 – 512 - 1)
Note: Negative values correspond to a trigger delay.
For <trigger mode> = IMMediate the value must be 0.
<# of samples>: Number of measurement values to record.
Value range: 1 to 130560 (= 128*1024 – 512)