TRIGger Subsystem
The TRIGger subsystem is used to synchronize instrument actions with events. It is thus possible to control
and synchronize the start of a sweep. An external trigger signal can be applied to the connector at the rear
panel of the instrument. A distinction is made between TRIGger1 (screen A) and TRIGger2 (screen B).
COMMAND PARAMETERS UNIT COMMENT
TRIGger<1|2>
[:SEQuence]
:SOURce
:LEVel
:AM
:FM
:PM
:IFPower
:RFPower
:VIDeo
:HOLDoff
[:TIME]
:ADJUST
:AUTO
:SLOPe
:VIDeo
:FORMat
:LPFRame
:FIELd
:SELect
:LINE
:NUMBer
:SSIGnal
:POLarity
:SYNChronize
:ADJust
:EXTernal
:IFPower
:IMMediate
:RFPower
IMMediate | EXTernal | VIDeo |
IFPower | RFPower TV | AF | AM |
FM | PM
<numeric_value>
<numeric_value>
<numeric_value>
<numeric_value>
<numeric_value>
<numeric_value>
<numeric_value>
POSitive|NEGative
525 | 625
ALL | ODD | EVEN
<numeric_value>
NEGative | POSitive
<numeric_value>
<numeric_value>
<numeric_value>
DBM
HZ
RAD
DBM
DBM
PCT
S
--
s
s
s
TV, RFPOWer with R&S FSP-B6
AF only with option R&S FS-K7
Option R&S FS-K7 FM Demodulator
Option R&S FS-K7 FM Demodulator
Option R&S FS-K7 FM Demodulator
Option R&S FSP-B6 TV- und RF-
Trigger
Option R&S FSP-B6 TV- and RF-
Trigger
Option R&S FSP-B6 TV- und RF-
Trigger
Option R&S FSP-B6 TV- und RF-
Trigger
Option R&S FSP-B6 TV- und RF-
Trigger
Option R&S FS-K5
TRIGger<1|2>[:SEQuence]:SOURce IMMediate | EXTernal | VIDeo | IFPower | RFPower | TV | AF |
FM | AM | PM
This command selects the trigger source for the start of a sweep.
Note: The selection RFPower or TV is only available with option R&S FSP-B6 (TV and RF
Trigger).
The selection AF, AM FM and PM is only available with option R&S FS-K7 (FM
Demodulator).
Parameter: IMMediate = automatic triggering the next measurement at the end of the
previous one.The value IMMediate corresponds to the FREE
RUN setting.
EXTernal = the next measurement is triggered by the signal at the external
trigger input.