Operating instructions

Status Reporting System R&S ESCI
1166.6004.12 5.24 E-1
STATus:QUEStionable Register
This register comprises information about indefinite states which may occur if the unit is operated
without meeting the specifications. It can be queried by commands STATus:QUEStionable:
CONDition? and STATus:QUEStionable[:EVENt]?.
Table 5-5 Meaning of bits in STATus:QUEStionable register
Bit No. Meaning
0 to 2 These bits are not used
3
POWer
This bit is set if a questionable power occurs (cf. also section "STATus:QUEStionable:POWer Register")
4
TEMPerature
This bit is set if a questionable temperature occurs.
5
FREQuency
The bit is set if a frequency is questionable (cf. section "STATus:QUEStionable:FREQuency Register")
6 to 7 These bits are not used
8
CALibration
The bit is set if a measurement is performed uncalibrated (=
^
label "UNCAL")
9
LIMit (device-specific)
This bit is set if a limit value is violated (see also section STATus:QUEStionable:LIMit Register)
10
LMARgin (device-specific)
This bit is set if a margin is violated (see also section STATus:QUEStionable:LMARgin Register)
11 This bit is not used
12
ACPLimit (device-specific)
This bit is set if a limit for the adjacent channel power measurement is violated (see also section
"STATus:QUEStionable:ACPLimit Register")
13
TRANsducer break (device-specific)
This bit is set if transducer break is reached
14 This bit is not used
15 This bit is always 0.