Operating instructions

R&S ESCI Status Reporting System
1166.6004.12 5.27 E-1
STATus:QUEStionable:LIMit<1|2> Register
This register comprises information about the observance of limit lines in the corresponding
measurement window (LIMit 1 corresponds to Screen A, LIMit 2 to Screen B). It can be queried with
commands STATus:QUEStionable:LIMit<1|2>:CONDition? and STATus:QUEStionable:
LIMit<1|2>[:EVENt]?.
Table 5-8 Meaning of bits in STATus:QUEStionable:LIMit<1|2> register
Bit No. Meaning
0
LIMit 1 FAIL
This bit is set if limit line 1 is violated.
1
LIMit 2 FAIL
This bit is set if limit line 2 is violated.
2
LIMit 3 FAIL
This bit is set if limit line 3 is violated.
3
LIMit 4 FAIL
This bit is set if limit line 4 is violated.
4
LIMit 5 FAIL
This bit is set if limit line 5 is violated.
5
LIMit 6 FAIL
This bit is set if limit line 6 is violated.
6
LIMit 7 FAIL
This bit is set if limit line 7 is violated.
7
LIMit 8 FAIL
This bit is set if limit line 8 is violated.
8 to 14 not used
15 This bit is always 0.