Operating instructions

R&S ESCI Status Reporting System
1166.6004.12 5.29 E-1
STATus:QUEStionable:POWer Register
This register comprises all information about possible overloads of the unit.
It can be queried with commands STATus:QUEStionable:POWer:CONDition? and "STATus
:QUEStionable:POWer[:EVENt]?.
Table 5-10 Meaning of bits in STATus:QUEStionable:POWer register
Bit No. Meaning
0
OVERload (Screen A)
This bit is set if the RF input is overloaded. 'OVLD' will then be displayed.
1
UNDerload (Screen A)
This bit is set if the RF input is underloaded. 'UNLD' will then be displayed.
2
IF_OVerload (Screen A)
This bit is set if the IF path is overloaded. 'IFOVL' will then be displayed.
3 to 7 not used
8
OVERload (Screen B)
This bit is set if the RF input is overloaded. 'OVLD' will then be displayed.
9
UNDerload (Screen B)
This bit is set if the RF input is underloaded. 'UNLD' will then be displayed.
10
IF_OVerload (Screen B)
This bit is set if the IF path is overloaded. 'IFOVL' will then be displayed.
11 to 14 not used
15 This bit is always 0.