Operating instructions

Status Reporting System R&S ESCI
1166.6004.12 5.30 E-1
STATus QUEStionable:TRANsducer Register
This register indicates that a transducer hold point is attained (bit 15) and what range is to be swept next
(bit 0 to 10). The sweep can be continued with command INITiate2:CONMeasure.
It can be queried with commands 'STATus:QUEStionable:TRANsducer:CONDition?' and
'STATus :QUEStionable:TRANsducer[:EVENt]?'.
Table 5-11 Meaning of bits in STATus:QUEStionable:TRANsducer register
Bit No. Meaning
0
Range 1
This bit is set when subrange 1 is attained.
1
Range 2
This bit is set when subrange 2 is attained.
2
Range 3
This bit is set when subrange 3 is attained.
3
Range 4
This bit is set when subrange 4 is attained.
4
Range 5
This bit is set when subrange 1 is attained.
5
Range 6
This bit is set when subrange 6 is attained.
6
Range 7
This bit is set when subrange 7 is attained.
7
Range 8
This bit is set when subrange 8 is attained.
8
Range 9
This bit is set when subrange 9 is attained.
9
Range 10
This bit is set when subrange 10 is attained.
10
not used
11
not used
12
not used
13
not used
14
Subrange limit
This bit is set when the transducer is at the point of changeover from one range to another.
15 This bit is always 0.