Operating instructions

R&S ESCI Error Messages
1166.6004.12 9.3 E-1
Continuation: Command Error
Error code Error text in the case of queue poll
Error explanation
-109
Missing parameter
The command contains too few parameters.
Example: The command SENSe:FREQuency:CENTer requires a frequency indication.
-110
Command header error
The header of the command is faulty.
-111
Header separator error
The header contains an imallowed separator.
Example: the header is not followed by a "White Space", "*ESE255"
-112
Program mnemonic too long
The header contains more than 12 characters.
-113
Undefined header
The header is not defined for the instrument.
Example: *XYZ is undefined for every instrument.
-114
Header suffix out of range
The header contains an imallowed numeric suffix.
Example: SENSe3 does not exist in the instrument.
-120
Numeric data error
The command contains a faulty numeric parameter.
-121
Invalid character in number
A number contains an invalid character.
Example: An "A" in a decimal number or a "9" in an octal number.
-123
Exponent too large
The absolute value of the exponent is greater than 32000.
-124
Too many digits
The number includes too many digits.
-128
Numeric data not allowed
The command includes a number which is not allowed at this position.
Example: The command INPut:COUPling requires indication of a text parameter.
-130
Suffix error
The command contains a faulty suffix.
-131
Invalid suffix
The suffix is invalid for this instrument.
Example: nHz is not defined.
-134
Suffix too long
The suffix contains more than 12 characters.
-138
Suffix not allowed
A suffix is not allowed for this command or at this position of the command.
Example: The command *RCL does not permit a suffix to be indicated.
-140
Character data error
The command contains a faulty text parameter
-141
Invalid character data
The text parameter either contains an invalid character or it is invalid for this command.
Example: Write error with parameter indication;INPut:COUPling XC.