Operating instructions

R&S ESCI Error Messages
1166.6004.12 9.5 E-1
Execution Error - Error on execution of a command; sets bit 4 in the ESR register
Error code Error text in the case of queue poll
Error explanation
-200
Execution error
Error on execution of the command.
-201
Invalid while in local
The command is not executable while the device is in local due to a hard local control.
Example: The device receives a command which would change the rotary knob state, but the device is
in local so the command can not be executed.
-202
Settings lost due to rtl
A setting associated with hard local control was lost when the device changed to LOCS from REMS or to
LWLS from RWLS.
-210
Trigger error
Error on triggering the device.
-211
Trigger ignored
The trigger (GET, *TRG or trigger signal) was ignored because of device timing considerations.
Example: The device was not ready to respond.
-212
Arm ignored
An arming signal was ignored by the device.
-213
Init ignored
Measurement initialisation was ignored as another measurement was already in progress.
-214
Trigger deadlock
The trigger source for the initiation of measurement is set to GET and subsequent measurement is
received. The measurement cannot be started until a GET is received, but the GET would cause an
interrupted-error)
-215
Arm deadlock
The trigger source for the initiation of measurement is set to GET and subsequent measurement is
received. The measurement cannot be started until a GET is received, but the GET would cause an
interrupted-error.
-220
Parameter error
The command contains a faulty or invalid parameter.
-221
Settings conflict
There is a conflict between setting of parameter value and instrument state.
-222
Data out of range
The parameter value lies out of the allowed range of the instrument.
-223
Too much data
The command contains too many data.
Example: The instrument does not have sufficient storage space.
-224
Illegal parameter value
The parameter value is invalid.
Example: The text parameter is invalid , TRIGger:SWEep:SOURce TASTe