Operating instructions

R&S ESCI Error Messages
1166.6004.12 9.7 E-1
Device Specific Error; sets bit 3 in the ESR register
Error code Error test in the case of queue poll
Error explanation
-300
Device-specific error
R&S ESCI-specific error not defined in greater detail.
-310
System error
This error message suggests an error within the instrument. Please inform the R&S Service.
-313
Calibration memory lost
Loss of the non-volatile data stored using the *CAL? command. This error occurs when the correction
data recording has failed.
-330
Self-test failed
The selftest could not be executed.
-350
Queue overflow
This error code is entered in the queue instead of the actual error code if the queue is full. It indicates
that an error has occurred but not been accepted. The queue can accept 5 entries.
Query Error - Error in data request; sets bit 2 in the ESR register
Error code Error text in the case of queue poll
Error explanation
-400
Query error
General error occurring when data are requested by a query.
-410
Query INTERRUPTED
The query has been interrupted.
Example: After a query, the instrument receives new data before the response has been sent completely.
-420
Query UNTERMINATED
The query is incomplete.
Example: The instrument is addressed as a talker and receives incomplete data.
-430
Query DEADLOCKED
The query cannot be processed.
Example: The input and output buffers are full, the instrument cannot continue operation.
-440
Query UNTERMINATED after indefinite response
A query is in the same command line after a query which requests an indefinite response.