Operating instructions

Error Messages R&S ESCI
1166.6004.12 9.8 E-1
Device-Specific Messages
Error code Error text in the case of queue poll
Error explanation
1036
MS: The correction table based amplifier gain exceeds the amplifier range for CALAMP1 and
CALAMP2 on IF board
This message is output when the setting range of the calibration amplifier is not sufficient for the required
correction. The error occurs only with modules which are not correcty adjusted or defective.
1052
Frontend LO is Unlocked
This message is output when the phase regulation of the local oscillator fails in the RF frontend.
1060
Trigger-Block Gate Delay Error- gate length < Gate Delay
This message is output when the gate signal length is not sufficient for the pull-in delay with a predefined
gate delay.
1064
Tracking LO is Unlocked
This message is output when the phase regulation of the local oscillator fails on the tracking generator
module.
Error code Error text in the case of queue poll
Error explanation
2022
OPTIONS.INI invalid
This message is output when an error has been recognized in the file OPTIONS.INI which contains the
clearing codes for retrofitable firmware applications. If this file is not correctly recognized, all firmware
applications are blocked for this instrument.
2028
Hardcopy not possible during measurement sequence
This message is output when a printout is started during scan sequences that cannot be interrupted.
Such sequences are for example:
Recording the system error correction data (calibration)
Instrument selftest
In such cases a synchronization to the end of the scan sequence should be performed prior to starting
the printout.
2033
Printer Not Available
This message is output when the selected printer is not included in the list of available output devices. A
possible cause is that the required printer driver is missing or incorrectly installed.
2034
CPU Temperature is too high
This message is output when the temperature of the processor exceeds 70 °C.