Operating instructions

ABORt - Subsystem R&S ESCI
1166.6004.12 6.8 E-1
ABORt Subsystem
The ABORt subsystem contains the commands for aborting triggered actions. An action can be
triggered again immediately after being aborted. All commands trigger events, and therefore they have
no *RST value.
COMMAND PARAMETERS UNIT COMMENT
:ABORt
:HOLD
--
--
--
--
no query
ABORt
This command aborts a current measurement and resets the trigger system.
Example: "ABOR;INIT:IMM"
Characteristics: *RST value: 0
SCPI: conforming
Mode: R, A, FM
:HOLD
This command interrupts a current scan measurement.
Example: "HOLD" 'interrupts the current scan measurement
Characteristics: *RST value: -
SCPI: conforming
Mode: R
CALCulate Subsystem
The CALCulate subsystem contains commands for converting instrument data, transforming and
carrying out corrections. These functions are carried out subsequent to data acquistion, i.e. following the
SENSe subsystem.
The numeric suffix is used in CALCulate to make the distinction between the two measurement
windows SCREEN A and SCREEN B:
CALCulate1 = Screen A
CALCulate2 = Screen B.
For commands without suffix, screen A is selected automatically.
Full Screen The settings are valid for the measurement window selected with the numeric
suffix. They become effective as soon as the corresponding measurement window
has been selected as active measurement window using the command
DISPLay[:WINDow<1|2>]:SELect. Triggering measurements and querying
measured values is possible only in the active measurement window.
Split Screen The settings are valid for the measurement window selected by means of the
numeric suffix and become effective immediately.
Note: In receiver mode, the markers can only be activated for the scan display. Therefore,
the numerical suffix is irrelevant in this operating mode.