Operating instructions

R&S ESCI CALCulate:DELTamarker Subsystem
1166.6004.12 6.9 E-1
CALCulate:DELTamarker Subsystem
The CALCulate:DELTamarker subsystem controls the delta-marker functions in the instrument. The
measurement windows are selected via CALCulate1 (screen A) or 2 (screen B).
COMMAND PARAMETERS UNIT COMMENT
CALCulate<1|2>
:DELTamarker<1 to 4>
[:STATe]
:MODE
:AOFF
:TRACe
:X
:RELative?
:Y?
:MAXimum
[:PEAK]
:NEXT
:RIGHt
:LEFT
:MINimum
[:PEAK]
:NEXT
:RIGHt
:LEFT
:LINK
<Boolean>
ABSolute|RELative
<numeric_value>
<numeric_value>
--
--
--
--
--
--
--
--
--
--
<Boolean>
--
--
HZ | S | DBM | DB
--
--
--
--
--
--
--
--
--
--
--
no query
query only
query only
no query
no query
no query
no query
no query
no query
no query
no query
:FUNCtion
:FIXed
[:STATe]
:RPOint
:Y
:OFFSet
:X
:PNOise
[:STATe]
:RESult?
<Boolean>
<numeric_value>
<numeric_value>
<numeric_value>
<Boolean>
--
DBM
DB
HZ | S
-- query only
CALCulate<1|2>:DELTamarker<1 to 4>[:STATe] ON | OFF
This command switches on and off the delta marker when delta marker 1 is selected. The
corresponding marker becomes the delta marker when delta marker 2 to 4 is selected. If the
corresponding marker is not activated, it will be activated and positioned on the maximum of the
measurement curve.
If no numeric suffix is indicated, delta marker 1 is selected automatically.
Example: "CALC:DELT3 ON" 'Switches marker 3 in screen A to delta marker mode.
Characteristics: *RST value: OFF
SCPI: device-specific
Mode: R, A, FM