CALCulate<1|2>:DELTamarker<1 to 4>:MAXimum:NEXT
This command positions the delta marker to the next smaller maximum value on the measured
curve. The corresponding delta marker will be activated first, if necessary.
Example: "CALC1:DELT2:MAX:NEXT" 'Sets delta marker 2 in screen A to the next
'smaller maximum value.
Characteristics: *RST value: -
SCPI: device-specific
Mode: R, A, FM
This command is an event and therefore has no *RST value and no query.
CALCulate<1|2>:DELTamarker<1 to 4>:MAXimum:RIGHt
This command positions the delta marker to the next smaller maximum value to the right of the
current value (i.e. ascending X values). The corresponding delta marker is activated first, if
necessary.
Example: "CALC2:DELT:MAX:RIGH" 'Sets delta marker 1 in screen B to the next
'smaller maximum value to the right of the
'current value.
Characteristics: *RST value: -
SCPI: device-specific
Mode: R, A, FM
This command is an event and therefore has no *RST value and no query.
CALCulate<1|2>:DELTamarker<1 to 4>:MAXimum:LEFT
This command positions the delta marker to the next smaller maximum value to the left of the
current value (i.e. descending X values). The corresponding delta marker will be activated first, if
necessary.
Example: "CALC:DELT:MAX:LEFT" 'Sets delta marker 1 in screen A to the next
'smaller maximum value to the left of the
'current value.
Characteristics: *RST value: -
SCPI: device-specific
Mode: R, A, FM
This command is an event and therefore has no *RST value and no query.
CALCulate<1|2>:DELTamarker<1 to 4>:MINimum[:PEAK]
This command positions the delta marker to the current minimum value on the measured crve. The
corresponding delta marker will be activated first, if necessary.
Example: "CALC2:DELT3:MIN" 'Sets delta marker 3 in screen B to the
'minimum value of the associated trace.
Characteristics: *RST value: -
SCPI: device-specific
Mode: R, A, FM
This command is an event and therefore has no *RST value and no query.