Operating instructions

R&S ESCI CALCulate:DELTamarker Subsystem
1166.6004.12 6.13 E-1
CALCulate<1|2>:DELTamarker<1 to 4>:MINimum:NEXT
This command positions the delta marker to the next higher minimum value of the measured curve.
The corresponding delta marker will be activated first, if necessary.
Example: "CALC1:DELT2:MIN:NEXT" 'Sets delta marker 2 in screen A to the next
higher minimum value.
Characteristics: *RST value: -
SCPI: device-specific
Mode: R, A, FM
This command is an event and therefore has no *RST value and no query.
CALCulate<1|2>:DELTamarker<1 to 4>:MINimum:RIGHt
This command positions the delta marker to the next higher minimum value to the right of the current
value (i.e. ascending X values). The corresponding delta marker will be activated first, if necessary.
Example: "CALC2:DELT:MIN:RIGH" 'Sets delta marker 1 in screen B to the next
'higher minimum value to the right of the
current value.
Characteristics: *RST value: -
SCPI: device-specific
Mode: R, A, FM
This command is an event and therefore has no *RST value and no query.
CALCulate<1|2>:DELTamarker<1 to 4>:MINimum:LEFT
This command positions the delta marker to the next higher minimum value to the left of the current
value (i.e. descending X values). The corresponding delta marker will be activated first, if necessary.
Example: "CALC:DELT:MIN:LEFT" 'Sets delta marker 1 in screen A to the next
'higher minimum to the left of the current
'value.
Characteristics: *RST value: -
SCPI: device-specific
Mode: R, A, FM
This command is an event and therefore has no *RST value and no query.
CALCulate<1|2>:DELTamarker<1>:LINK ON | OFF
This command links delta marker 1 to marker 1.The x-value of delta marker 1 is kept fixed at 0.
Therefore, delta marker 1 always has the same x-value as marker 1. Changing the x-value of marker
1 automatically changes the value of delta marker 1. This function is useful if marker 1 is positioned
on trace 1 and delta marker 2 on trace 2.
For DELTamarker only numeric suffix 1 is permitted.
Example: "CALC:DELT:LINK ON" 'links delta marker 1 to marker 1
Characteristics: *RST value: -
SCPI: device-specific.
Mode:R, A