Operating instructions

R&S ESCI CALCulate:LIMit Subsystem
1166.6004.12 6.19 E-1
COMMAND PARAMETERS UNIT COMMENT
CALCulate<1|2>
:LIMit<1 to 8>
:TRACe
:STATe
:UNIT
:FAIL?
:CLEar
[:IMMediate]
:COMMent
:COPY
:NAME
:DELete
<numeric_value>
<Boolean>
DBM | DBPW | WATT |
DBUV | DBMV | VOLT |
DBUA | AMPere | DBPT |DB |
DBUV_M | DBUA_M
--
<string>
1 to 8 | < name>
<string>
--
--
--
--
--
--
query only
no query
CALCulate<1|2>:LIMit<1 to 8>:TRACe 1 to 3
This command assigns a limit line to a trace in the indicated measurement window.
Examples: "CALC:LIM2:TRAC 3" 'Assigns limit line 2 to trace 3 in screen A.
"CALC2:LIM2:TRAC 1" 'Assigns limit line 2 to trace 1 in screen B at
the same time.
Characteristics: *RST value: 1
SCPI: device-specific
Mode: R, A, FM
CALCulate<1|2>:LIMit<1 to 8>:STATe ON | OFF
This command switches on or off the limit check for the selected limit line in the selected
measurement window. The result of the limit check can be queried with CALCulate:LIMit<1 to
8>:FAIL?.
Example: "CALC:LIM:STAT ON" 'Switches on the limit check for limit line 1 in
'screen A.
"CALC2:LIM:STAT OFF" 'Switches off the limit check for limit line 1 in
'screen B.
Characteristics: *RST value: OFF
SCPI: conforming
Mode: R, A, FM
CALCulate<1|2>:LIMit<1 to 8>:UNIT DBM | DBPW | WATT | DBUV | DBMV | VOLT |DBUA | AMPere |
DB | DBT | DBUV_M | DBMV_M
This command defines the unit of the selected limit line. The definition is valid independently of the
measurement window.
Upon selection of the unit DB the limit line is automatically switched to the relative mode. For units
different from DB the limit line is automatically switched to absolute mode.
Example: "CALC:LIM4:UNIT DBUV" 'Sets the unit of limit line 4 to dBยตV.
Characteristics: *RST value: DBM
SCPI: device-specific
Mode: R, A, FM
In receiver mode the units DBM, DBUV, DBUA, DBPW, DBPT, DBUV_M, and DBUA_M are
available
In analyzer mode all units are available except for DBPT.