Operating instructions

CALCulate:LIMit Subsystem R&S ESCI
1166.6004.12 6.24 E-1
CALCulate<1|2>:LIMit<1 to 8>:ACPower:ACHannel:ABSolute:STATe ON | OFF
This command activates the limit check for the adjacent channel when adjacent-channel power
measurement (Adjacent Channel Power) is performed. Before the command, the limit check for the
channel/adjacent-channel measurement must be globally switched on using CALC:LIM:ACP ON.
The result can be queried with CALCulate:LIMit:ACPower:ACHannel:RESult?. It should be
noted that a complete measurement must be performed between switching on the limit check and
the result query, since otherwise no valid results are available.
The numeric suffixes <1 to 8> in LIMIt are irrelevant for this command.
Example: "CALC:LIM:ACP:ACH 30DB, 30DB"
'Sets the relative limit value in screen A for
'the power in the lower and upper adjacent
'channel to 30 dB below the channel power.
"CALC:LIM:ACP:ACH:ABS -35DBM, -35DBM"
'Sets the absolute limit value in screen A
'for the power in the lower and upper
'adjacent channel to -35 dBm.
"CALC:LIM:ACP ON" 'Switches on globally the limit check for the
'channel/adjacent channel measurement in
'screen A.
"CALC:LIM:ACP:ACH:REL:STAT ON"
'Switches on the check of the relative limit
'values for adjacent channels in screen A.
"CALC:LIM:ACP:ACH:ABS:STAT ON"
'Switches on the check of absolute limit 'values
for the adjacent channels in screen 'A.
"INIT;*WAI" 'Starts a new measurement and waits for
'the sweep end.
"CALC:LIM:ACP:ACH:RES?" 'Queries the limit check result in the
'adjacent channels in screen A.
Characteristics: *RST value: OFF
SCPI: device-specific
Mode: A