Operating instructions

CALCulate:LIMit Subsystem R&S ESCI
1166.6004.12 6.28 E-1
CALCulate<1|2>:LIMit<1 to 8>:ACPower:ALTernate<1|2>:ABSolute -200DBM to 200DBM,
-200DBM to .200DBM
This command defines the absolute limit value for the lower/upper alternate adjacent channel power
measurement (Adjacent Channel Power) in the selected measurement window.
The numeric suffix after ALTernate<1|2> denotes the first or the second alternate channel. The
numeric suffixes <1 to 8> are irrelevant for this command.
It should be noted that the absolute limit value for the limit check has no effect as soon as it is below
the relative limit value defined with CALCulate:LIMit:ACPower:ALTernate<1|2>:RELative.
This mechanism allows automatic checking of the absolute basic values defined in mobile radio
standards for the power in adjacent channels.
Parameter: The first value is the limit for the lower and the upper alternate adjacent channel.
The second limit value is ignored but must be indicated for reasons of
compatibility with the FSE family.
Example: "CALC:LIM:ACP:ALT2:ABS -35DBM, -35DBM"
'Sets the absolute limit value in
'screen A for the power in the lower
'and upper second alternate
'adjacent channel to -35 dBm.
Characteristics: *RST value: -200DBM
SCPI: device-specific
Mode: A