1166.6004.12 6.30 E-1
CALCulate<1|2>:LIMit<1 to 8>:ACPower:ALTernate<1|2>:RESult?
This command queries the result of the limit check for the first/second alternate adjacent channel in
the selected measurement window for adjacent channel power measurements.
The numeric suffix after ALTernate<1|2> denotes the first or the second alternate channel. The
numeric suffixes <1 to 8> are irrelevant for this command.
If the power measurement of the adjacent channel is switched off, the command produces a query
error.
Parameter: The result is returned in the form <result>, <result> where
<result> = PASSED | FAILED and where the first (second) returned value
denotes the lower (upper) alternate adjacent channel.
Example: "CALC:LIM:ACP:ALT2 30DB, 30DB"
'Sets the relative limit value in screen A for the
'power in the lower and upper second alternate
'adjacent channel to 30 dB below the channel
'power.
"CALC:LIM:ACP:ALT2:ABS -35DBM, -35DBM"
'Sets the absolute limit value in screen A for the
'power in the lower and upper second alternate
'adjacent channel to -35 dBm.
"CALC:LIM:ACP ON" 'Switches on globally the limit check for the
'channel/adjacent channel measurement in
'screen A.
"CALC:LIM:ACP:ALT:STAT ON"
'Switches on the limit check for the adjacent
'channels in screen A.
"INIT;*WAI" 'Starts a new measurement and waits for the
'sweep end.
"CALC:LIM:ACP:ALT:RES?" 'Queries the limit check result in the second
'alternate adjacent channels in screen A.
Characteristics: *RST value: --
SCPI: device-specific
Mode: A
This command is a query and therefore has no *RST value.