Operating instructions

R&S ESCI CALCulate:LIMit Subsystem
1166.6004.12 6.31 E-1
CALCulate:LIMit:CONTrol Subsystem
The CALCulate:LIMit:CONTrol subsystem defines the x axis (CONTrol-axis).
COMMAND PARAMETERS UNIT COMMENT
CALCulate<1|2>
:LIMit<1...8>
:CONTrol
[:DATA]
:DOMain
:OFFSet
:MODE
:SHIFt
:SPACing
<numeric_value>,<numeric_value>..
FREQuency|TIME
<numeric_value>
RELative|ABSolute
<numeric_value>
LINear | LOGarithmic
HZ | S
HZ | S
HZ | S
CALCulate<1|2>:LIMit<1 to 8>:CONTrol[:DATA] <numeric_value>,<numeric_value>..
This command defines the X axis values (frequencies or times) of the upper or lower limit lines.The
values are defined independently of the measurement window.
The number of values for the CONTrol axis and for the corresponding UPPer and/or LOWer limit
lines have to be identical. Otherwise default values are entered for missing values or not required
values are deleted.
In analyzer mode, the unit of values depends on the frequency or time domain of the X axis, i.e. it is
HZ with CALC:LIM:CONT:DOM FREQ und S bei CALC:LIM:CONT:DOM TIME.
Example: "CALC:LIM2:CONT 1MHz,30MHz,100MHz, 300MHz,1GHz"
'Defines 5 reference values for the X axis of
'limit line 2
"CALC:LIM2:CONT?" 'Outputs the reference values for the X axis of
'limit line 2 separated by a comma.
Characteristics: *RST value: - (LIMit:STATe is set to OFF)
SCPI: conforming
Mode: R, A, FM
CALCulate<1|2>:LIMit<1 to 8>:CONTrol:DOMain FREQuency | TIME
This command defines the frequency or time domain for the x axis values.
Example: "CALC:LIM2:CONT:DOM TIME" 'Defines the time domain for the X axis of limit
'line 2.
Characteristics: *RST value: FREQuency
SCPI: device-specific
Mode: A, FM