Operating instructions

R&S ESCI CALCulate:LIMit Subsystem
1166.6004.12 6.33 E-1
CALCulate:LIMit:LOWer Subsystem
The CALCulate:LIMit:LOWer subsystem defines the lower limit line. This subsystem is not available in
receiver mode.
COMMAND PARAMETERS UNIT COMMENT
CALCulate<1|2>
:LIMit<1 to 8>
:LOWer
[:DATA]
:STATe
:OFFSet
:MARGin
:MODE
:SHIFt
:SPACing
:THReshold
<numeric_value>,<numeric_value>..
<Boolean>
<numeric_value>
<numeric_value>
RELative|ABSolute
<numeric_value>
LINear | LOGarithmic
<numeric_value>
DBM | DB |
DEG | RAD |
S | HZ | PCT
--
DB| DEG| RAD|
S | HZ | PCT
DB| DEG| RAD|
S | HZ | PCT
--
DB| DEG| RAD|
S | HZ | PCT
DBM | DB |
DEG | RAD |
S | HZ | PCT
CALCulate<1|2>:LIMit<1 to 8>:LOWer[:DATA] <numeric_value>,<numeric_value>..
This command defines the values for the selected lower limit line independently of the measurement
window.
The number of values for the CONTrol axis and for the corresponding LOWer limit line has to be
identical. Otherwise default values are entered for missing values or not necessary values are
deleted.
The unit must be identical with the unit selected by CALC:LIM:UNIT. If no unit is indicated, the unit
defined with CALC:LIM:UNIT is automatically used.
If the measured values are smaller than the LOWer limit line, the limit check signals errors.
Example: "CALC:LIM2:LOW -30,-40,-10,-40,-30"
'Defines 5 lower limit values for limit line 2 in
'the preset unit.
"CALC:LIM2:LOW?" 'Outputs the lower limit values of limit line 2
'separated by a comma.
Characteristics: *RST value: - (LIMit:STATe is set to OFF)
SCPI: conforming
Mode: A, FM