Operating instructions

CALCulate:LIMit Subsystem R&S ESCI
1166.6004.12 6.36 E-1
CALCulate:LIMit:UPPer Subsystem
The CALCulate:LIMit:UPPer subsystem defines the upper limit line.
COMMAND PARAMETERS UNIT COMMENT
CALCulate<1|2>
:LIMit<1 to 8>
:UPPer
[:DATA]
:STATe
:OFFSet
:MARGin
:MODE
:SPACing
:SHIFt
:SPACing
:THReshold
<numeric_value>,<numeric_value>..
<Boolean>
<numeric_value>
<numeric_value>
RELative|ABSolute
LINear | LOGarithmic
<numeric_value>
LINear | LOGarithmic
<numeric_value>
DBM | DB
--
DB
DB
--
--
DB
DBM | DB
CALCulate<1|2>:LIMit<1 to 8>:UPPer[:DATA] <numeric_value>,<numeric_value>..
This command defines the values for the upper limit lines independently of the measurement
window.
The number of values for the CONTrol axis and for the corresponding UPPer and/or LOWer limit line
have to be identical. Otherwise default values are entered for missing values or not necessary values
are deleted.
The unit must be identical with the unit selected by CALC:LIM:UNIT. If no unit is indicated, the unit
defined with CALC:LIM:UNIT is automatically used.
In analyzer mode, the limit check signals errors if the measured values exceed the UPPer limit line.
Example: "CALC:LIM2:UPP -10,0,0,-10,-5" 'Defines 5 upper limit values for limit
'line 2 in the preset unit.
"CALC:LIM2:UPP?" 'Outputs the upper limit values for limit line 2
'separated by a comma.
Characteristics: *RST value: - (LIMit:STATe is set to OFF)
SCPI: conforming
Mode: R, A, FM