Operating instructions

R&S ESCI CALCulate:LIMit Subsystem
1166.6004.12 6.37 E-1
CALCulate<1|2>:LIMit<1 to 8>:UPPer:STATe ON | OFF
This command switches on or off the indicated limit line in the selected measurement window. The
limit check is activated separately with CALC:LIM:STAT ON.
In analyzer mode, the result of the limit check can be queried with CALCulate:LIMit<1 to
8>:FAIL?.
Example: "CALC1:LIM4:UPP:STAT ON" 'Switches on limit line 4 (upper limit) in
'screen A.
"CALC2:LIM4:UPP:STAT ON" 'Switches on limit line 4 (upper limit) in
'screen B.
Characteristics: *RST value: OFF
SCPI: conforming
Mode: R, A, FM
CALCulate<1|2>:LIMit<1 to 8>:UPPer:OFFSet <numeric_value>
This command defines an offset for the Y axis of the selected relative upper limit line. In contrast to
CALC:LIM:UPP:SHIFt, the line is not shifted by modifying the individual Y values but by means of
an additive offset. The offset is independent of the measurement window.
Example: "CALC:LIM2:UPP:OFFS 3dB" 'Shifts limit line 2 by 3 dB upwards in the
'corresponding measurement windows.
Characteristics: *RST value: 0
SCPI: device-specific
Mode: A, FM
CALCulate<1|2>:LIMit<1 to 8>:UPPer:MARGin <numeric_value>
This command defines a margin to an upper limit line, at which out-of-limit values are signalled (if the
limit check is active), but not handled as a violation of the limit value. The margin is independent of
the measurement window.
Example: "CALC:LIM2:UPP:MARG 10dB" 'Defines the margin of limit line 2 to 10 dB
'below the limit value.
Characteristics: *RST value: 0
SCPI: device-specific
Mode: R, A, FM