Operating instructions

R&S ESCI CALCulate:MARKer Subsystem
1166.6004.12 6.39 E-1
CALCulate:MARKer Subsystem
The CALCulate:MARKer subsystem checks the marker functions in the instrument. The measurement
windows are assigned to CALCulate 1 (screen A) or 2 (screen B).
COMMAND PARAMETERS UNIT COMMENT
CALCulate<1|2>
:MARKer<1 to 4>
[:STATe]
:AOFF
:TRACe
:X
:SLIMits
[:STATe]
:LEFT
:RIGHt
:COUNt
:RESolution
:FREQuency?
:COUPled
[:STATe]
:SCOupled
[:STATe]
:LOEXclude
:Y?
:PERCent
:MAXimum
[:PEAK]
:NEXT
:RIGHt
:LEFT
:MINimum
[:PEAK]
:NEXT
:RIGHt
:LEFT
:PEXCursion
<Boolean>
<numeric_value>
<numeric_value>
<Boolean>
<numeric_value>
<numeric_value>
<Boolean>
<numeric_value>
--
<Boolean>
<Boolean>
<Boolean>
--
<numeric_value>
--
--
--
--
--
--
--
--
<numeric_value>
--
--
HZ | S | DBM | DB
HZ | S
HZ | S
--
HZ
--
--
PCT
--
--
--
--
--
--
--
--
--
DB
no query
query only
query only
no query
no query
no query
no query
no query
no query
no query
no query
CALCulate<1|2>:MARKer<1 to 4>[:STATe] ON | OFF
This command switches on or off the currently selected marker in the selected measurement
window. If no indication is made, marker 1 is selected automatically. If marker 2, 3 or 4 is selected
and used as a delta marker, it is switched to marker mode.
Example: "CALC:MARK3 ON" 'Switches marker 3 in screen A on or to marker mode.
Characteristics: *RST value: OFF
SCPI: device-specific
Mode: R, A, FM