Operating instructions

CALCulate:MARKer Subsystem R&S ESCI
1166.6004.12 6.40 E-1
CALCulate<1|2>:MARKer<1 to 4>:AOFF
This command switches off all active markers in the selected measurement window. All delta
markers and active marker/delta marker measurement functions are switched off.
Example: "CALC:MARK:AOFF" 'Switches off all markers in screen A.
Characteristics: *RST value: -
SCPI: device-specific
Mode: R, A, FM
This command is an event and therefore has no *RST value and no query.
CALCulate<1|2>:MARKer<1 to 4>:TRACe 1 to 3
This command assigns the selected marker (1 to 4) to the indicated measurement curve in the
selected measurement window. The corresponding trace must be active, i.e. its status must be
different from "BLANK".
If necessary the corresponding marker is switched on prior to the assignment.
Example: "CALC:MARK3:TRAC 2" 'Assigns marker 3 in screen A to trace 2.
"CALC2:MARK:TRAC 3" 'Assigns marker 1 in screen B to trace 3.
Characteristics: *RST value -
SCPI: device-specific
Mode: R, A, FM
CALCulate<1|2>:MARKer<1 to 4>:X 0 to MAX (frequency | sweep time)
This command positions the selected marker to the indicated frequency (span > 0), time (span = 0) or level
(APD measurement or CCDF measurement ON) in the selected measurement window. If marker 2, 3 or
4 is selected and used as delta marker, it is switched to marker mode.
Available units: receiver: Hz; analyzer: Hz
Example: "CALC1:MARK2:X 10.7MHz" 'Positions marker 2 in screen A to frequency
'10.7 MHz.
Characteristics: *RST value: -
SCPI: device-specific
Mode: R, A, FM
CALCulate<1|2>:MARKer<1 to 4>:X:SLIMits[:STATe] ON | OFF
This command switches between a limited (ON) and unlimited (OFF) search range in the selected
measurement window. The function is independent of the selection of a marker, i.e. the numeric
suffix MARKer<1 to 4> is irrelevant.
If the time domain power measurement is active, this command limits the evaluation range on the
trace.
Example: "CALC:MARK:X:SLIM ON" 'Switches on search limitation in screen A.
Characteristics: *RST value: OFF
SCPI: device-specific
Mode: R, A, FM