Operating instructions

R&S ESCI CALCulate:MARKer Subsystem
1166.6004.12 6.41 E-1
CALCulate<1|2>:MARKer<1 to 4>:X:SLIMits:LEFT 0 to MAX (frequency | sweep time)
This command sets the left limit of the search range for markers and delta markers in the selected
measurement window. Depending on the x axis domain the indicated value defines a frequency
(span > 0) or time (span = 0). The function is independent of the selection of a marker, i.e. the
numeric suffix in MARKer<1 to 4> is irrelevant.
If the time domain power measurement is active, this command limits the evaluation range to the
trace.
Note: The function is only available if the search limit for marker and delta marker is switched on
(CALC:MARK:X:SLIM ON).
Example: "CALC:MARK:X:SLIM ON" 'Switches the search limit function on
for screen A.
"CALC:MARK:X:SLIM:LEFT 10MHz" 'Sets the left limit of the search range in
screen A to 10 MHz.
Characteristics: *RST value: - (is set to the left diagram border on switching on search limits)
SCPI: device-specific
Mode: R, A, FM
CALCulate<1|2>:MARKer<1 to 4>:X:SLIMits:RIGHT 0 to MAX (frequency | sweep time)
This command sets the right limit of the search range for markers and delta markers in the selected
measurement window. Depending on the x axis domain the indicated value defines a frequency
(span > 0) or time (span = 0). The function is independent of the selection of a marker, i.e. the
numeric suffix in MARKer<1 to 4> is irrelevant.
If the time domain power measurement is active, this command limits the evaluation range to the
trace.
Note:
The function is only available if the search limit for marker and delta marker is switched on
(CALC:MARK:X:SLIM ON).
Example: "CALC:MARK:X:SLIM ON" 'Switches the search limit function on
'for screen A.
"CALC:MARK:X:SLIM:RIGH 20MHz" 'Sets the right limit of the search range
'in screen A to 20 MHz.
Characteristics: *RST value: - is set to the right diagram border on switching on search
limits)
SCPI: device-specific
Mode: R, A, FM